Urinalysis
Dipstick Interpretation
Cheryl S. Sine, DVM;
Paula Krimer, DVM, DVSc; Perry J. Bain, DVM, PhD; and Kenneth S. Latimer,
DVM, PhD
Class of 2003 (Sine)
and Department of Pathology (Krimer, Bain, Latimer), College of Veterinary
Medicine, The University of Georgia, Athens, GA 30602-7388

Introduction
Urinalysis is an
important tool in disease detection, as well as monitoring and screening
animal health. Abnormalities can be indicative of diseases of the urinary
system as well as other organ systems, including liver function, acid-base
status, and carbohydrate metabolism.1 Complete urinalysis
involves both macroscopic and microscopic assessment. This is typically
performed by gross visual assessment of the urine, microscopic examination,
and chemical evaluation. Several chemical parameters can be measured
using a commercially available in house dipstick test. This test is
relatively inexpensive, and takes less than 5 minutes to complete.
Typical dipstick strips include the following tests: bilirubin, blood,
glucose, ketones, pH, protein specific gravity, and urobilinogen. Some
dipsticks also include leukocytes and nitrite analyses.
 |
| Figure
1: Example of commercially available Bayer® reagent
strips for urinalysis. |
Sample
Collection
Urine should be
collected in a clean, dry container that is free of any disinfecting
or cleaning chemicals. Samples may be collected by free catch of
voided sample, manual bladder expression, catheterization, or cystocentesis.2
Voided samples
are the easiest and least invasive samples to collect. However, voided
samples may have contaminants that include bacteria, epithelial cells,
and white blood cells.1 Red blood cells should not be
found in normal voided samples. Voided samples should be collected
midstream to lessen contaminants from the vagina or prepuce.3,4 Collection
of samples from surfaces such as floors, cages, and litter boxes
should be avoided, since these will introduce environmental contaminants.
Manual expression
of the bladder is another technique used in urine collection. In
this method, the patients bladder is gently squeezed until
urine is expressed. This technique may lead to bladder trauma resulting
in hematuria, and in some instances (such as urethral obstruction)
may result in a ruptured bladder.5 This method may have
the same cellular contaminants as a voided sample.
Catheterization
is performed by placing a small hollow tube into the urethra to the
level of the bladder. Urine is then withdrawn from the bladder using
a syringe. Catheterized samples have less contamination from the
distal urogenital tract; however, contamination from the urethra
may still occur. Contaminants include epithelial cells or red blood
cells. Poor catheterization technique may lead to trauma or, less
commonly, infection.3,4,5
Cystocentesis samples
are collected by inserting a sterile needle through the body wall
into the bladder. Urine is withdrawn from the bladder using a syringe.
A lateral or ventral approach to the bladder may be made without
causing severe trauma to any vital region of the bladder. Clipping
or surgical preparation of the area along the body wall is not necessary
prior to sample collection. Often a 1 inch or 1.5 inch 22 gauge needle
is used attached to a 6 or 12 cc syringe. The bladder is manually
immobilized and the needle is inserted through the abdominal wall
into the bladder, and the urine is withdrawn. It is important to
stop aspirating prior to withdrawing the needle as this may lead
to aspiration of blood cells or epithelium from the bladder wall.
Animals often tolerate cystocentesis very well and little restraint
is needed. Contaminants that may be found include iatrogenically
introduced red blood cells. 3,4,5 Rarely, enterocentesis
may occur which results in a sample containing bacteria, intestinal
villi and other intestinal contents.
Sample
Handling
In order to obtain
accurate results, the urine collection, storage and handling must
be sterile and follow standard procedures. The dipstick analysis
should be performed as soon after collection as possible (ideally
within 30 minutes of collection) and the sample should be well mixed
prior to testing. If for some reason the test cannot be performed
immediately, the sample may be covered and refrigerated. It should
be allowed to return to room temperature prior to testing. The dipsticks
should be stored in the original airtight container to maintain reagent
reactivity.5,6
Testing
Methods
Dipsticks may be
removed from the air tight, light sealed containers. It is important
not to touch the reagent areas of the strip as this may alter test
results. Each reagent area should be immersed in urine by dipping.
The excess urine should be removed to prevent dilution of reagents
or mixing of reagents between pads. This can be achieved by tilting
the strip and allowing the urine to run off the edges (Figure 2).
While blotting excess urine, ensure the chemicals from the different
tests do not mix.6
 |
| Figure
2: Removal of excess urine horizontally will prevent
the mixing of chemicals from different reagent pads. |
The reagent pads
should be read at the specified times. These times are different
for each test and also vary between dipstick manufacturers. Compare
the blocks to the corresponding color chart provided by the test
strip's manufacturer.6
Urine discoloration
may create difficulty in visually interpreting the test results.
Color changes may be masked, or read as false positive test results.
If the urine is noticeably discolored, the sample may be centrifuged
and the supernatant used for analysis. Common causes of altered urine
color include hemoglobin, myoglobin, bilirubin, and drugs (phosphoenolpyruvate).
Interpretation
of Test Results
Dipsticks
commonly include tests for specific gravity, pH, glucose, protein,
blood, bilirubin, ketones, urobilinogen, nitrite, and leukocytes.
Specific
Gravity
Urine specific
gravity is based on the ratio of weight of urine to weight of an
equivalent volume of pure water. This test is used to measure tubular
function. The dipstick measures specific gravity by measuring the
change in pKa of polyelectrolytes in relation to ionic concentration5.
Although dipstick
strips do have a method of approximating specific gravity, this measurement
is best made with a refractometer.1
Urine specific
gravity measured by the dipstick can be falsely elevated by moderate
to high concentrations of protein. Low reading may occur if the urine
is alkaline.6 High lipid content in urine may also alter
the results by either raising or lowering the specific gravity measurement.
pH
The pH of urine
can vary depending on an animals diet as well as its acid-base
status. For example, animals that primarily eat high protein meat-based
diets will have acidic urine. On the other hand, animals that eat
more vegetable-based diets will have an alkaline urine.1,5
Dipsticks measure
pH using methyl red, bromthymol blue or phenolphathalein indicator
dyes. These reagents react rapidly and result in a color change.
pH can be approximated within 0.5 units using a dipstick. The urine
sample should be fresh as urine becomes more alkaline on standing
due to the conversion of urea to ammonia by bacteria (if present),
and loss of CO2.1
Causes of acidic
urine include: meat diet, systemic acidosis, hypochloridemia, and
administration of acidifying agents such as d,l-methionine or NH4Cl. 4,5 Urine
with high concentrations of glucose may have a lower pH.6 This
is due to bacterial metabolism of glucose and and production of ammonia
which lowers pH.1
Causes of alkaline
urine include: vegetable based diet, bacterial infection of urease-producing
bacteria, systemic alkalosis, urine exposed to room air for an extended
time (loss of CO2), and administration of alkalinizing agents including
citrate or NaHCO3.1,4,5
Urine pH also may
provide good predictive assessment of crystal and stone morphology
as certain crystals and stones form in either acidic or alkaline
environments. Uric acid, cystine, and calcium oxalate crystals are
found in acidic urine. On the other hand, struvite, calcium carbonate,
calcium phosphate, ammonium biurate, and amorphous phosphate crystals
are found in alkaline urine.5
For a more accurate
assessment of urine pH, a pH meter may be used. However, for most
routine veterinary analyses a dipstick pH reading is sufficient.5
Glucose
Glucose is not
detectable in the urine of healthy dogs or cats. In a healthy animal,
glucose passes freely through the glomerular filter and is resorbed
by the proximal tubules. If glucosuria is present, it is due to either
an excess amount of glucose reaching the tubules that cannot be resorbed
or, less commonly, decreased tubular resorptive function.4
Reagent strips
measure glucose levels using the glucose oxidase method.1 This
method is a sequential enzymatic reaction. Glucose reacts with glucose
oxidase to produce hydrogen peroxide, which oxidizes the indicator
chemical to produce a color change. The color change is related to
the amount of glucose present in the urine sample (Figure 3).
 |
| Figure
3: Example of marked glucosuria with Bayer® reagent
strips. |
Glucosuria may
be either persistent or transient and multiple tests may be needed
for differentiation of these conditions. Persistent causes of glucosuria
include: diabetes mellitus, administration of glucose containing
fluids, chronic disease that is not related to the kidneys such as
hyperadrenocorticism, hyperpituitarism, or acromegaly. Other diseases
that may result in transient hyperglycemias leading to glucosuria
include: hyperthyroidism, acute pancreatitis, stress (especially
in cats), postprandial, and administration of certain drugs. Rarely,
a Fanconi-like syndrome may lead to glucosuria.1,4
False positive
test results may be caused by contamination of the sample with oxidants
such as hydrogen peroxide, bleach (sodium hypochlorite), or occasionally
pseudoglucose in obstructed cats.1,5
False negative
test results may be due to high concentrations of ascorbic acid (Vitamin
C) in the urine. Moderately high concentrations of ketones also may
cause false negative test results if the amount of glucose is only
slightly elevated. The glucose test also becomes less reactive as
urine specific gravity increases or as temperature decreases. Cold
urine (refrigerated specimens) or expired reagent strips may also
result in false negative test results.1,5,6
Protein
Dogs and cats normally
have small proteins that pass through the glomerular filter, however
a majority of these proteins are resorbed by the renal tubules. The
renal nephron does excrete a small amount of Tamm Horsfall protein.
Thus, only a very small amount of protein is normally excreted in
the urine, which is not usually clinically detectable.1,5
The protein portion
of the dipstick reagent strip measures the protein based on a pH
dye indicator method using bromphenol blue. Due to the negative charge
of albumin, if protein (albumin) is present in urine, the pH increases,
and a positive test result occurs. This test is primarily sensitive
to albumin is relatively insensitive for the detection of globulins
and Bence-Jones proteins (see Figure 4).1
 |
| Figure
4: Example of marked proteinuria with Bayer® reagent
strips. |
Positive protein
results must be evaluated in relationship to the patients history,
physical examination, method of urine collection, urine specific
gravity, and microscopic sediment examination. Proteinuria may be
due to hemorrhage, infection, intravascular hemolysis, or renal disease.
Hemorrhage is confirmed by a positive occult blood reaction on the
dipstick and the presence of red blood cells in the sediment. A urinary
infection or cystitis can be confirmed by observing bacteria and
white blood cells on sediment examination. Cases of intravascular
hemolysis have hemoglobinuria leading to a positive occult blood
test.1,4,5
Proteinuria of
renal disease may be due to glomerular and/or tubular lesions. If
the proteinuria is due to renal disease, the occult blood test will
be negative and the sediment may or may not contain casts. Determination
of the urine protein/urine creatinine ratio is helpful in confirming
renal proteinuria.1
Protein results
must be analyzed with the urine specific gravity. Trace proteinuria
may represent significant protein loss with low specific gravity,
but not with high specific gravity.1,4
False positive
protein reactions may occur with alkaline urine or if a disinfectant
residue is in the urine, possibly from improper cleaning of the collection
container.6 Samples containing urease-producing bacteria
may have an elevated pH resulting in a false positive test result.
False negative
test results may occur in dilute or acidic urine.6
If the urine protein
dipstick is positive for protein, the sample should be further analyzed
with a quantitative method at an outside laboratory.
Blood
The occult blood
test will react positively in the presence of red blood cells, free
hemoglobin or free myoglobin (see Figure 5). Hemoglobin usually is
bound and is too large to pass through the glomerular filter. If
the renal threshold is exceeded, the hemoglobin can pass into the
urine. Myoglobin on the other hand, is not bound and freely passes
through the glomerular filter. Myoglobin can be detected in urine
before a change in plasma color is apparent. The presence of free
red blood cells results in a positive test when blood cells lyse
and hemoglobin is released. Healthy animals should have negative
test results.1,5
This test is based
on a pseudoperoxidase reaction, which is more sensitive to hemoglobin
and myoglobin than intact red blood cells.5 This test
is also more sensitive to hemoglobin than the urine protein dipstick
tests.1
 |
| Figure
5: Example of trace hematuria and marked hematuria or
hemoglobinuria with Bayer® reagent strips. |
A positive occult
blood test indicates hematuria, hemoglobinuria, or myoglobinuria.
Further evaluation of the urine sediment is needed if a positive
test result is found. Most commonly, hematuria is the cause of the
positive test result while myoglobinuria is rare.1
Hematuria can be
caused by trauma, infection, inflammation, infarction, calculi, neoplasia
or a coagulopathy anywhere along the urinary tract.5 In
cases of hematuria, the urine is red and cloudy, but will clear if
centrifuged. Microscopic evaluation of the urine sediment will reveal
red blood cells.1
Hemoglobinuria,
on the other hand, will have reddish brown urine that does not become
clear after centrifugation. The microscopic evaluation of urine sediment
will not reveal red blood cells. With intravascular hemolysis, plasma
will have a reddish tint due to hemoglobinemia that is detectable
prior to hemoglobinuria. The patient usually will be clinically anemic.1
A false positive
test result may occur if the urine is contaminated with bleach, or
contains large amounts of iodide or bromide. If a voided sample is
collected from a bitch in heat, a false positive test may also occur.
In this case a cystocentesis sample is preferred for analysis. Microbial
peroxidase that is present in some urinary tract infections, can
also lead to false positive test results.5,6
False negative
test results may occur if the urine is not well mixed prior to evaluation.
This is due to the fact that red blood cells often sediment quickly.5
Bilirubin
Bilirubin is produced
from the breakdown of hemoglobin, transported to the liver bound
to albumin, and conjugated with carbohydrates by hepatocytes. Only
conjugated bilirubin is found in urine. Excess bilirubin may be produced
when red blood cells are destroyed, or in liver disease, including
bile duct obstruction. Conjugated bilirubin is detected in urine
if the renal threshold is exceeded. The renal threshold in dogs,
especially males, is lower than that of other species.1
Reagent strips
measure levels of conjugated bilirubin with the diazotization method.
This occurs by coupling bilirubin with diazotized dichloroaniline
in an acidic environment.6 Bilirubin is very unstable
when exposed to room air and light. Thus, urine specimens should
be tested soon after collection (see Figure 6).6
 |
| Figure
6: Example of marked bilirubinuria with Bayer® reagent
strips. |
Positive test results
may be observed in concentrated urine of healthy dogs. In dogs, the
renal threshold for bilirubin is low and renal tubules are able to
break down heme and produce some renal bilirubin, therefore slight
bilirubinuria can be a normal finding in dogs with concentrated urine.4 However,
bilirubinuria is always abnormal in cats. Bilirubinuria may indicate:
liver disease, bile duct obstruction, starvation, hemolysis, or pyrexia.
Bilirubinuria in bile duct obstruction is often more severe than
that of hepatocellular disease.1,5
False positive
test results may occur if high doses of chlorpromazine, which lowers
urine pH, have been given. A metabolite of etodolac (Lodine), also
produces false positive test results.6
False negative
test results may occur in urine samples with high ascorbic acid or
nitrite concentration.5
Ketones
Acetone, acetoacetic
acid, and beta-hydroxybutyric acid are ketones. Glomeruli freely
filter ketones and the tubules then resorb them completely. If the
tubular resorptive capacity is saturated, then the ketones are incompletely
resorbed, resulting in ketonuria. Ketonuria occurs quickly in younger
animals and is more easily detected than ketonemia. Ketonuria does
not signify renal disease, but rather excessive lipid or defective
carbohydrate metabolism.1
Dipstick tests
are semiquantitative and only detect acetone and acetoacetic acid.
Reagent strips contain nitroprusside that does not react with beta-hydroxybutyric
acid.1,5
Ketonuria may be
caused by starvation, insulinoma, diabetic ketoacidosis, persistent
hypoglycemia, high fat low carbohydrate diets, and glycogen storage
disease1,5.
False positive
test results may occur if urine is pigmented, or has high concentrations
of levodopa metabolites.5,6
False negative
test results are uncommon in fresh urine, but may occur if urine
samples are old. Ketones are very volatile and can evaporate quickly.5,6
Urobilinogen
Urobilinogen is
formed by intestinal bacteria from the breakdown of conjugated bilirubin.
Urobilinogen is usually excreted in feces, however a small amount
may be reabsorbed and excreted in urine. This test is not of significant
value in animals.1,5
The dipstick method
measures urobilinogen by reacting with p-diethylaminobenzaldehyde
in an acid environment.6
A positive test
response indicates normal enterohepatic circulation of biliary pigments.
High concentrations of biliary pigments may occur in hemolytic crisis,
or cases of hepatic or intestinal dysfunction.1
A false positive
test result may occur if the temperature of the reagent strip is
elevated.6
A false negative
test result may occur if there is formalin residue in the collection
container, or if the sample is old, because urobilinogen is very
unstable when exposed to light and air.6
Nitrite
The nitrite portion
of the dipstick analysis has limited value in veterinary medicine.
This is due to the high number of false negative test results in
small animals. Nitrites occur in urine during some bacterial infections.
In order to achieve an accurate positive test result, the urine must
have been retained in the bladder at least 4 hours. Therefore, it
is best to collect a (first) morning sample or ensure the patient
has not urinated in at least 4 hours.1,5
A positive test
indicates a bacterial infection. Gram negative rods are more likely
produce a positive test response.6
Negative test results
do not exclude infection. The urinary tract infection may involve
organisms that do not convert nitrites, or the urine may not have
been held in the bladder greater than 4 hours.6
Leukocytes
The leukocyte test
detects the presence of white blood cells or partial cells in the
urine. In dogs, this test is indicative of pyuria but false negative
test results often occur.5
Leukocytes are
measured by a reaction of the esterases in leukocytes that catalyze
reaction of pyrrole amino acid ester to release 3-hydroxy-5-phenol
pyrrole.5,6
False positive
test results often occur in cats, and this test is clinically unreliable.5 False
positive test results also may occur in the event of fecal contamination.
False negative
test results may develop if the patient has been treated with high
doses of tetracycline or other antibiotics. Glucosuria or increased
urine specific gravity may cause false negative test results.6 False
negative test results may be observed with voided urine samples obtained
from animals with pyometra or prostatitis.
Conclusion
The dipstick portion
of urinalysis is an important diagnostic and monitoring laboratory
test system. There are certain techniques of sample collection, handling
and testing that should be followed closely to ensure accurate results.
The causes of false positive and negative test results also should
be considered when evaluating urine dipstick results. Because the
dipstick test is easy to perform and economical, all veterinary practices
should be able to perform this test in house rather than submitting
samples to an outside laboratory. In house laboratory analysis also
provides more accurate results and prevents age-induced artifacts
that may lead to false positive and false negative test results.
Table
1. Potential causes of error when evaluating urine
via dipstick method5
- Refrigerated
urine sample not returned to room temperature prior to testing
- Urine
contaminated with disinfectant from skin or cleaning
prior to collection
- Expired
reagent strips
- Improper
storage with exposure to air or light
- Leakage
of reagent chemical from one test to another if the test
is read vertically rather than horizontally
- Tests
read at inappropriate times
- Highly
pigmented urine
- Failure
to use control urine to check accuracy of strip
|
Table
2. Summary of Dipstick Analysis
| Parameter |
Expected
Results |
Interpretation of
Results |
Causes
of False Positives (or Increase) |
Causes
of False Negatives (or Decrease) |
| Specific
Gravity |
1.005
to 1.065 |
|
moderate
to high levels of protein |
alkaline
urine |
| pH |
5.0
to 6.0 |
Acidic:
meat diet; acidosis; low chloride; acidifying agents Alkaline: vegetable
based diet; bacterial infection; alkalosis; urine exposed to
air for extended times; administration of alkalinizing agents;
postprandial tide |
glucose
in urine |
|
| Glucose |
Negative
to Trace |
Positive:
Chronic or transient hyperglycemia; post administration of certain
drugs; rarely-fanconi-like syndrome |
hydrogen
peroxide; bleach |
ascorbic
acid; ketones; increased specific gravity; cold urine; expired
reagent strips |
| Protein |
Negative
to Trace |
hemorrhage;
urinary infection; intravascular hemolysis; renal disease |
alkaline
urine; disinfectant residue |
dilute
or acidic urine |
| Blood |
Negative |
Positive:
hematuria: trauma, infection, inflammation, infarction, calculi,
neoplasia, or coagulopathy hemoglobinuria:
myoglobinuria |
bleach
contamination; high levels of bromide or iodide; bitch in heat |
poorly
mixed urine |
| Bilirubin |
Negative |
Positive:
can be normal in dogs; ALWAYS abnormal in cats; indicates liver
disease, bile duct obstruction, starvation; hemolysis; pyrexia |
high
doses of chlorpromazine; etodolac metabolites |
ascorbic
acid; nitrates |
| Ketones |
Negative |
Positive:
starvation; insulinoma; diabetes mellitus; persistent hypoglycemia;
high fat low carbohydrate diets; glycogen storage disease |
pigmented
urine |
old
urine sample |
| Urobilinogen |
0.2
to 1.0 mg/dL |
Positive:
hemolytic crisis; intestinal or hepatic dysfunction |
elevated
reagent strip temperature |
old
urine sample, formalin residue in collection container |
| Nitrite |
Negative |
bacterial
infection |
|
common
in small animals |
| Leukocytes |
Negative |
pyuria |
common in cats,
and if fecal contamination |
often occurs;
tetracycline; high glucose; high specific gravity; voided sample
in animal with pyometra or prostatitis |
Bibliography
1. Duncan, J.R.,
Prasse, K.W., & Mahaffey, E.A.. Veterinary Laboratory Medicine:
Clinical Pathology.3rd ed. Iowa State University Press,
Ames, 1994. pp 162-183.
2. Ling, G.V.. Techniques
of urine collection and handleing. In: Ling G.V.Lower Urinary Tract
Diseases of Dogs and Cats. St. Louis, Mosby, 1995. pp 23-28.
3. Zinkl, J.G. in
Cowell, R.L., Tyler, R.D., & Meinkoth, J.H.. Urinary Sediment
and Cytology of the Urinary Tract. in Diagnostic Cytology and
Hematology of the Dog and Cat. 2nd ed. Mosby, Inc, St. Louis,
1999. pp 211-229.
4. Osborne, C.A.,
Low, D.G., & Finco, D.R.. Canine and Feline Urology. W.B.Saunders
Co., Philadelphia, 1972. pp 25-61.
5. Chew, D.J & DiBartola,
S.P.. Interpretation of Canine and Feline Urinalysis. Ralston
Purina Co., St Louis, 1998. pp1-21.
6. Package insert
for BAYER Multistix® 10SG · Multistix® 9 · Multistix® 9
SG · Multistix® 8 SG · Multistix® 7 · N-Multistix® SG · Multistix® 9 · N-Multistix® · Multistix® SG · Multistix® · Bili-Labstix® Reagent
Strips. Bayer Corporation, Elkhart, 1992.
7. Package Insert
fro Ictotest® Reagent Tablets for Urinalysis. Bayer Corporation,
Elkhart, 1995.
8. Package Insert
for Acetest® Reagent Tablets. Miles, Inc, 1980 |